PHYSICAL SPECIFICATION
• Dimensions & Weight: Height Width Length Weight
Analyzer 42 cm 36 cm 32cm 14 Kg
16.5 in 14.2 in 12.6 in 30.9 lb
• Printer: 80 column dot matrix printer
• Throughput:
- Up to 60 samples/hour in Open Tube
- Up to 55 samples/hour in Close Tube
• Sound Pressure Level: <60 dBa
• Operating Temperature & Humidity:
16 - 34°C (61 - 93°F) room temperature
Maximum relative humidity 80% for temperature up to 31°C (88°F)
decreasing linearity humidity at 40°C (104°F)
• Specimen Volume: CBC Mode - 10 μl / CBC 3 part DIFF - 10μl
• Power Requirements:
Power supply from 90 Vac to 240 Vac
50 Hz to 60 Hz
Power consumption Maximum 230 VA
• Reagents: 3 reagents only (ABX minipack or bottles):
- ABX Minidil or ABX Minidil LMG
- ABX Minilyse or ABX Minilyse LMG or ABX Lysebio (cyanide free)
- ABX Miniclean
METHODS & TECHNOLOGIES
• RBC/PLT detection principles
Method Impedance
Ruby diameter 50 μm
Depression of count 200 mb
Duration of count 2 (or 3) x 6 seconds
Dilution ratio 1/15 000
• HGB measurement
Method Photometry
Wavelenght 555 nm
Dilution ratio 1/250
• HCT measurement
Method Numeric integration
• WBC measurement
Method Impedance
Ruby diameter 80 μm
Depression of count 200 mb
Duration of count 2 (or 3) x 6 seconds
Dilution ratio 1/250
• 3 part Differentiation
Method Impedance + specific lyse action
• MCV,MCH,MCHC,RDW,PCT*,PDW* Calculation
NORMES
• CE UL CSA
NF EN 60601-1. UL 3101-1 C22.2 n°1010-1
NF EN 61326. B
NF EN 61000-3-2
NF EN 61000-3-3
SOFTWARE SPECIFICATIONS
• Data Processing :
- LCD screen with 2 lines of 40 alphanumerical characters
- Capacity: Last result + graphics or 78 results
with optional memory smart card
` - Mother board: 68HC11 microprocessor
- Transmit results & QC to LIS
- Unidirectional connection RS 232
• Quality control management :
- Optional smart card based QC (31 controls x 3 levels)