Specifications
CLIA Classification : Moderate complexity
Sample Type : Whole blood
Sample Size : 100 μL Syringe 75 μL Capillary
Measured Parameters
pH 6.900 – 7.900 pH units
PCO2 8.0 – 150.0 mmHg
PO2 10 – 700 mmHg
Calculated Parameters
pH (T)
(pH temperature corrected)
PCO2 (T)
(PCO2 temperature corrected)
PO2 (T)
(PO2 temperature corrected)
TCO2 0 – 50 mmol/L
(Total Carbon Dioxide)
HCO3- 0 – 50 mmol/L
(Bicarbonate)
BEb -25.0 – 25.0 mmol/L
(Base Excess in blood)
BEecf -25.0 – 25.0 mmol/L
(Base Excess in extra cellular fluid)
SBC 0 – 50 mmol/L
(Standard Bicarbonate)
%SO2C 40.0 – 100.0%
(Oxygen Saturation calculated at normal P50)
A-aDO2 0 – 700 mmHg
(Alveolar arterial oxygen gradient)
RI 0.0 – 70.0
(Respiratory Index)
Input Parameters
Patient Temperature 5 – 45°C
Hemoglobin 30 – 30.0 g/dL
FIO2 10 – 100%
Patient ID 14 digits
Sample Temperature Control: 37°C +–
0.2°C
Ambient Conditions
: 15 – 30°C (59 – 86°F), 500 – 800 mmHg (max 15 PSI)
5 – 85% relative humidity, non-condensing
atmospheric air environment (21% PO2)
Analysis Time: 125 seconds
Data Storage
: 64 Patient results with Operator ID, Patient ID, Date and Time
QC – up to 30 results for each Level (1, 2, 3)
Calibration
: Automatic or On-Demand
Input/Output
: Numeric keypad, graphic display, 27 column thermal printer,
barcode reader port, RS-232 computer interface port
Power : 100/115 ~ VAC, 50 – 60 Hz, 0.8 A or 220 ~ VAC, 50 – 60 Hz, 0.4A
Size & Weight : 14.5" W x 12.5" H x 7" D (37 cm W x 32 cm H x 18 cm D)
16 lbs. (7.3 kg) with reagent module